Imbert-Fick law

The Imbert-Fick "law" was invented by Hans Goldmann (1899–1991) to give his newly marketed tonometer (with the help of the Haag-Streit Company) a quasi-scientific basis; it is mentioned in the ophthalmic and optometric literature, but not in any books of physics. According to Goldmann,[1] "The law states that the pressure in a sphere filled with liquid and surrounded by an infinitely thin membrane is measured by the counterpressure which just flattens the membrane." "The law presupposes that the membrane is without thickness and without rigidity...practically without any extensibility."

The problem is that a sphere formed by such a membrane and filled with incompressible liquid (water) cannot be indented or applanated even when the pressure inside is zero, because a sphere contains the maximum volume with the minimum surface area.[2][3] Any deformation necessarily increases surface area, which is impossible if the membrane is inelastic, such as aluminum foil.

The actual physical basis of tonometry is Newton's third law of motion: "If you press a stone eyeball with your finger, the finger is also pressed by the stone eyeball."

The law is this:

Intraocular pressure = Contact force/Area of contact

The law assumes that the cornea is infinitely thin, perfectly elastic, and perfectly flexible.[4] None of these assumptions are true. The cornea is a membrane that has thickness and offers resistance when pressed.[5] Therefore, in Goldmann tonometry, readings are normally taken when an area of 3.06mm diameter has been flattened. At this point the opposing forces of corneal rigidity and the tear film are roughly approximate in a normal cornea and cancel each other out allowing the pressure in the eye to be inferred from the force applied.[6]

Notes

  1. Goldmann H. "Applanation Tonometry". Transactions Second Glaucoma Conference. New York, Josiah Macy, Jr Foundation. 1957.
  2. Koster W. "Beiträge zur Tonometrie und Manometrie des Auges". Graefe's Arch. Ophthalmol. 1895; 41: 113-158.
  3. Markiewitz HH. "The so-called Imbert-Fick Law". AMA Arch. Ophthalmol. 1960; 64: 189/159.
  4. Whitacre, MM, Stein, R. (1993) "Sources of error with use of Goldmann-type tonometers". Surv. Ophthalmol. 38,1-30
  5. Anders Eklund, Per Hallberg, Christina Lindén, and Olof A. Lindahl. (2003) "An Applanation Resonator Sensor for Measuring Intraocular Pressure Using Combined Continuous Force and Area Measurement". Invest. Ophthalmol. Vis. Sci. 2003 Jul;44(7):3017-24.
  6. The Glaucoma Book, Paul N. Schacknow, John R. Samples, p.79. Springer, 2010. ISBN 978-0-387-76699-7.
This article is issued from Wikipedia - version of the 6/11/2016. The text is available under the Creative Commons Attribution/Share Alike but additional terms may apply for the media files.